Research Method
Uncertainty evaluation for the determination of 10,11-dihydro-10-hydroxycarbazepine in human plasma by UHPLC-MS/MS
Full
Wei LI1, Xue-qiang PAN1, Zhi-gang ZHAO2, Sheng-hui MEI2
Affiliations
  • 1.Department of Pharmacy, Beijing Health Vocational College, Beijing 101101, China
  • 2.Department of Pharmacy, Beijing Tiantan Hospital, Capital Medical University, Beijing 100070, China
Published: 2025-12-17 doi: 10.13699/j.cnki.1001-6821.2025.23.010
Outline
收藏切换
Objective

To evaluate the uncertainty for the determination of 10,11-dihydro-10-hydroxycarbazepine (MHD) in human plasma by ultra performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS).

Methods

An isotope-labeled internal standard was employed. Chromatographic separation was achieved within 3 min on a BEH C18 column (2.1 mm×50.0 mm, 1.7 μm) using a gradient elution at a flow rate of 0.4 mL·min-1, with a mobile phase consisting of methanol containing 0.1% formic acid and water containing 5% methanol plus 0.1% formic acid. Plasma sample concentrations were determined by UHPLC-MS/MS with electrospray ionization (ESI) in multiple reaction monitoring (MRM) mode. The method was validated for selectivity, calibration curve, lower limit of quantification, precision, recovery, matrix effect, and stability. The influencing factors in the quantification of MHD were analyzed and evaluated to calculate the individual uncertainties, combined uncertainty, and finally the expanded uncertainty.

Results

MHD demonstrated good linearity within the range of 0.03 to 6.00 mg·L-1. The accuracy (bias from the theoretical concentration) for within-run and between-run analyses was within ±15%, and the precision (coefficient of variation) was within 13.32%. Plasma samples showed good stability under the following conditions: 9 hours at 25 ℃, 5 hours at 4 ℃, three freeze-thaw cycles, and storage at -80 ℃ for 18 days. Based on the calculated uncertainty values for each contributing factor, repeatability, matrix effect, and recovery represented the major sources of uncertainty for the low-concentration quality control (QC) sample, with values of 1.96×10-2, 5.49×10-2, and 7.25×10-2, respectively. For the high-concentration QC sample, matrix effect and recovery contributed substantially to the overall uncertainty, with values of 4.92×10-2 and 2.01×10-2, respectively. The expanded uncertainties (P=95%, k=2) for MHD at low (0.09 mg·L-1) and high (4.50 mg·L-1) concentrations in human plasma were 1.62×102 mg·L-1 and 0.48 mg·L-1, respectively.

Conclusion

The measurement uncertainty of MHD concentration in human plasma determined by UHPLC-MS/MS is primarily contributed by repeatability, matrix effect and recovery at low concentration levels, while dominated by matrix effect and recovery at high concentrations

10,11-dihydro-10-hydroxycarbazepine  /  ultra performance liquid chromatography-tandem mass spectrometry  /  human plasma  /  uncertainty
Wei LI, Xue-qiang PAN, Zhi-gang ZHAO, Sheng-hui MEI. Uncertainty evaluation for the determination of 10,11-dihydro-10-hydroxycarbazepine in human plasma by UHPLC-MS/MS[J]. Chinese Journal of Clinical Pharmacology, 2025 , 41 (23) : 3366 -3373 . DOI: 10.13699/j.cnki.1001-6821.2025.23.010
Year 2025 volume 41 Issue 23
PDF
142
30
Cite this Article
BibTeX
Article Info
doi: 10.13699/j.cnki.1001-6821.2025.23.010
  • Receive Date:2025-07-03
  • Online Date:2026-08-05
  • Published:2025-12-17
Article Data
Affiliations
History
  • Received:2025-07-03
Funding
Affiliations
    1.Department of Pharmacy, Beijing Health Vocational College, Beijing 101101, China
    2.Department of Pharmacy, Beijing Tiantan Hospital, Capital Medical University, Beijing 100070, China
References
Share
https://castjournals.cast.org.cn/joweb/zglcylxzz/EN/10.13699/j.cnki.1001-6821.2025.23.010
Share to
QR

Scan QR to access full text

Cite this article
BibTeX
Citations
表12种不同金属材料的力学参数

Family
属数
Number of
genus
种数
Number of
species
占总种数比例
Percentage of
total species (%)

Genus
种数
Number of
species
占总种数比例
Percentage of total
species (%)
鹅膏菌科Amanitaceae 2 11 5.26 鹅膏菌属 Amanita 10 4.78
小菇科 Mycenaceae 2 12 5.74 丝盖伞属 Inocybe 5 2.39
多孔菌科 Polyporaceae 8 14 6.70 蜡蘑属 Laccaria 5 2.39
红菇科 Russulaceae 3 23 11.00 小皮伞属 Marasmius 6 2.87
小菇属 Mycena 11 5.26
光柄菇属 Pluteus 5 2.39
红菇属 Russula 17 8.13
栓菌属 Trametes 5 2.39
关闭全屏
  • BibTeX
  • EndNote
  • RefWorks
  • TxT