Performance quality control testing and analysis of ventilators
Full
Ji Zhang, Jianwei Qi*
Affiliations
  • Zhejiang Hospital, Hangzhou 310013, China
Published: 2025-10-10 doi: 10.11876/mimt202505007
Outline
收藏切换
Objective

To evaluate the performance of 450 ventilators in use in Zhejiang Province by systematically assessing the pass rates of core parameters across different brands and analyzing the correlation between service life and performance, thereby providing evidence for precise quality control and scientific maintenance.

Methods

In accordance with GB 9706.212—2020 and JJF 1234—2018, five key parameters (tidal volume, mandatory ventilation frequency, FiO2, inspiratory pressure level, and PEEP) were tested using a Fluke VT650 gas flow analyzer. Descriptive statistics and Spearman rank correlation analysis were employed for systematic comparison and correlation analysis from the dimensions of brand type and service life.

Results

The overall pass rate of the 450 ventilators was 58.67% (264/450). Significant disparities were observed in the pass rates of core parameters among brands: Imported Brand 2 had the highest overall pass rate (70.48%), while Domestic Brand 2 had the lowest (0.00%). Tidal volume exhibited the most notable variation in pass rates among brands (0%~90.75%), identified as the primary weakness of domestic devices. Correlation analysis between service life and performance revealed that Imported Brand 2 maintained stable performance with no significant correlation to service life (ρ=-0.094, P=0.157). In contrast, Imported Brand 1 (ρ=-0.382, P=0.001) and Imported Brand 3 (ρ=-0.446, P<0.001) showed significant negative correlations, while Domestic Brand 1 also demonstrated a weak negative correlation (ρ=-0.215, P=0.035).

Conclusion

The current overall performance of ventilators requires improvement, with tidal volume accuracy being a common technical bottleneck for domestic devices. Furthermore, performance degradation patterns over time vary significantly by brand. It is recommended to establish a differentiated quality control system focusing on tidal volume, incorporating service life and brand-specific characteristics, to enhance the clinical safety and whole-life-cycle management of ventilators.

Ventilator  /  Quality control  /  Statistical analysis  /  Service years  /  Tidal volume
张吉, 祁建伟. 呼吸机性能质量控制检测与分析. 现代仪器与医疗, 2025 , 31 (5) : 48 -52 . DOI: 10.11876/mimt202505007
Ji Zhang, Jianwei Qi. Performance quality control testing and analysis of ventilators[J]. Modern Instruments & Medical Treatment, 2025 , 31 (5) : 48 -52 . DOI: 10.11876/mimt202505007
Year 2025 volume 31 Issue 5
48
10
Cite this Article
BibTeX
Article Info
doi: 10.11876/mimt202505007
  • Online Date:2026-10-23
  • Published:2025-10-10
Article Data
Affiliations
History
Funding
Affiliations
    Zhejiang Hospital, Hangzhou 310013, China
References
Share
https://castjournals.cast.org.cn/joweb/xdyqyyl/EN/10.11876/mimt202505007
Share to
QR

Scan QR to access full text

Cite this article
BibTeX
Citations
表12种不同金属材料的力学参数
科
Family
属数
Number of
genus
种数
Number of
species
占总种数比例
Percentage of
total species (%)
属
Genus
种数
Number of
species
占总种数比例
Percentage of total
species (%)
鹅膏菌科Amanitaceae 2 11 5.26 鹅膏菌属 Amanita 10 4.78
小菇科 Mycenaceae 2 12 5.74 丝盖伞属 Inocybe 5 2.39
多孔菌科 Polyporaceae 8 14 6.70 蜡蘑属 Laccaria 5 2.39
红菇科 Russulaceae 3 23 11.00 小皮伞属 Marasmius 6 2.87
小菇属 Mycena 11 5.26
光柄菇属 Pluteus 5 2.39
红菇属 Russula 17 8.13
栓菌属 Trametes 5 2.39
关闭全屏
  • BibTeX
  • EndNote
  • RefWorks
  • TxT