收藏切换
Evaluation of measurement uncertainty of calibration results of temperature deviation in high and low temperature test chamber
收藏切换
PDF
Ting-Ting ZHAO*, Zhao-Liang ZHOU
Laboratory Testing | 2024, 2(7) : 136 - 140
Less
收藏切换
Laboratory Testing | 2024, 2(7): 136-140
Evaluation and Analysis
Evaluation of measurement uncertainty of calibration results of temperature deviation in high and low temperature test chamber
Full
Ting-Ting ZHAO*, Zhao-Liang ZHOU
Affiliations
  • Xi'an Thermal Power Research Institute Co., Ltd. Xi'an 710032 China
Outline
收藏切换

The high and low temperature test chamber can provide thecorresponding high and low temperature environment for instruments andmaterials, and its temperature field performance and measurementuncertainty are important indicators that affect the measurement resultsof the temperature test chamber. Through the uncertainty evaluation andinfluence analysis of temperature deviation calibration results, themeasurement performance of high and low temperature test chamber, errorsources and methods to reduce the influence of error are furtherunderstood. According to the calibration method specified in GB/T10592-2023 Technical Conditions for High and Low Temperature TestChamber, select the temperature sensor that meets the measurementrequirements and calibrate the temperature deviation of the high and lowtemperature test chamber. The calibration confirms that the temperaturefield performance of the high and low temperature test chamber meets thetolerance requirement of temperature deviation $\pm {2}^{\circ }\mathrm{C}$ , and themeasurement uncertainty meets the technical index requirement of $1/3$ maximum allowable error. Thetemperature field performance of the high-low test chamber is good, andthe short-branch A-class platinum thermal resistance selected by themeasurement standard meets the calibration requirements and has goodstability.

high and low temperature test chamber  /  temperature deviation  /  calibration  /  measurement uncertainty
Ting-Ting ZHAO, Zhao-Liang ZHOU. Evaluation of measurement uncertainty of calibration results of temperature deviation in high and low temperature test chamber[J]. Laboratory Testing, 2024 , 2 (7) : 136 -140 .
Year 2024 volume 2 Issue 7
PDF
11
5
Cite this Article
BibTeX
Article Info
  • Online Date:2025-07-29
Article Data
Affiliations
History
Affiliations
    Xi'an Thermal Power Research Institute Co., Ltd. Xi'an 710032 China

Corresponding:

*ZHAO Ting-Ting, Engineer, Xi’an Thermal Power Research Institute Co., Ltd., No. 136 Xingqing Road, Xi’an 710032, China. E-mail:
References
Share
https://castjournals.cast.org.cn/joweb/sysjc/EN/
Share to
QR

Scan QR to access full text

Cite this article
BibTeX
Citations
表12种不同金属材料的力学参数

Family
属数
Number of
genus
种数
Number of
species
占总种数比例
Percentage of
total species (%)

Genus
种数
Number of
species
占总种数比例
Percentage of total
species (%)
鹅膏菌科Amanitaceae 2 11 5.26 鹅膏菌属 Amanita 10 4.78
小菇科 Mycenaceae 2 12 5.74 丝盖伞属 Inocybe 5 2.39
多孔菌科 Polyporaceae 8 14 6.70 蜡蘑属 Laccaria 5 2.39
红菇科 Russulaceae 3 23 11.00 小皮伞属 Marasmius 6 2.87
小菇属 Mycena 11 5.26
光柄菇属 Pluteus 5 2.39
红菇属 Russula 17 8.13
栓菌属 Trametes 5 2.39
关闭全屏
  • BibTeX
  • EndNote
  • RefWorks
  • TxT