Latest ArticlesA liquid-liquid extraction-high performance liquid chromatography triple quadrupole mass spectrometry(HPLC-MS/MS) method was established to analyze the residual levels of four aflatoxins(AF), namely AFB1, AFB2, AFG1, and AFG2, in the fermented drugs. The sample was dissolved in water and extracted with dichloromethane, and then concentrated by nitrogen blowing and dissolved in 30% methanol. Finally, it was detected by the HPLC-MS/MS method. The results showed that it was linear for four AFs in the range of 0.10 - 10 ng/L, with a detection limit of 0.03 μg/kg and a quantification limit of 0.1 μg/kg. In the fermented drug matrix, the spiked recovery rate of AFs was 80% - 105%, with the RSD of 0.2% - 5%. The sample testing results showed that AF was not detected in any of the 8 fermented drugs.The established method is simple, sensitive, and highly accurate, which can quantitatively detect AF and provide strong technical support for the quality control of fermented drugs and AF contamination control.
The detection methods were established for diethyl sulfate(2) and ethyl hydrogen sulphate(3) as the genotoxic impurities in sildenafil citrate(1). Compound 2 was determined by GC-MS method with TG-5SILMS capillary column(30 m×0.25 mm×0.25 μm) as the chromatographic column, helium as the carrier gas, and selected ion monitoring(SIM) mode for determination. Compound 3 was determined by LC-MS/MS method with an Agilent Zorbax C18-AQ Plus column (4.6 mm×100 mm, 3.5 μm). And the analysis was carried out in the gradient elution mode with 0.1% formic acid solution containing 50 mmol/L ammonium formate as mobile phase A, and methanol as mobile phase B. The electrospray ionization(ESI) source and multi reaction monitoring(MRM) mode were adopted. The results showed that it was linear for 2 and 3 in the ranges of 21.1 - 422.6 and 4.06 - 81.2 ng/mL, respectively. The average recovery rates(n=9) of 2 and 3 were 104.6% and 104.8%, respectively, with detection limits of 7.04 and 2.03 ng/mL, and quantification limits of 21.1 and 4.06 ng/mL, respectively. The established method is accurate, sensitive, and convenient, which can be used for the determination of 2 and 3 in 1.
An inductively coupled plasma mass spectrometry(ICP-MS) method was established to determine 8 elemental impurities, namely Pb, V, As, Hg, Cd, Co, Ni, and Cr, in tenofovir alafenamide fumarate(1). The sample was dissolved and diluted with an internal standard solution(containing Bi, Y, In, and Sc elements). Then, the sample was directly injected into the ICP-MS system. The results showed that it was linear for the 8 elements in the corresponding concentration ranges. The detection limits were 0.21 - 7.69 μg/kg, and the quantification limits were 0.71 - 25.60 μg/kg. The sample recovery rates(n=9) were 93.41% - 108.88%, the RSDs(n=6) of the repeatability test were 0.64% - 1.56%. The solution had good stability within 2 hours at room temperature. The content of each element impurity in multiple batches of samples was far below its limit value. The established method has good specificity, high sensitivity, and good accuracy and precision, which can meet the quality control of element impurities in 1.
A headspace gas chromatography(HS-GC) method was established to determine the contents of four impurities, namely acetaldehyde, formaldehyde, ethylene oxide, and dioxane, in octylphenol polyoxyethylene ether 40(OP-40). The standard addition method was employed for the analysis. The sample was dispersed with water as the solvent, and equilibrate at 70 ℃ for 30 min in the headspace chamber. A hydrogen flame ionization detector(FID) was used, and a capillary column(60 m×0.25 mm×1.4 μm) was employed for measurement at programmed temperature. The results showed that under the proposed chromatographic conditions, the 4 compounds were well separated, and it was linear for the four impurities in the range of 0.05 - 21.56 μg/g. The detection limits were 0.02 - 0.32 μg/g. The average recovery rates(n=9)of acetaldehyde, formaldehyde, ethylene oxide, and dioxane were 103.3%, 99.8%, 99.1% and 100.1%, respectively.The established method is applicable for the daily detection of four impurities in OP-40, which provides a reference for improving its quality specification
Rheological methods exhibit significant potential in characterizing the microstructure of semisolid preparations. However, there is a lack of fully established standardized procedures for this technique. The objective of this study was to develop and validate rheological methods, aiming to provide a reference for establishing standardized protocols. Taking fluticasone furoate nasal spray as a model, this study assessed the impacts of the set parameters for the measurement methods in flow sweep, flow ramp, oscillation amplitude, and oscillation frequency modes on the selected key rheological parameters(yield value, infinite shear viscosity, thixotropic loop area, critical stress, storage modulus, loss modulus, and loss tangent). The results of methodological investigations(repeatability, precision, specificity, and robustness) showed that the RSD values of many measured parameters exceeded 15%. Hence, the determination of equivalent standard ranges for parameters such as rheological microstructure required further studies.
Fluticasone furoate nasal spray is used to treat allergic rhinitis, and its efficacy and patient experience are significantly influenced by the rheological properties, including yield value, thixotropy, linear viscoelastic region, and maximum modulus. These characteristics not only affect the distribution and retention time of the drug within the nasal cavity, but also impact bioavailability and overall therapeutic effectiveness of the product. This study aimed to conduct an in-depth analysis of the rheological properties of fluticasone furoate nasal spray. Then, the differences in these characteristics between the self-prepared formulation and the commercially available preparation were determined and assessed using a rheometer. The self-prepared formulation exhibited a broader linear viscoelastic region compared with the commercial product, suggesting superior uniformity and stability. The maximum modulus value of the self-prepared formulation was lower than that of the commercial product, indicating that the self-prepared formulation was more easily deformed and better distributed in the nasal cavity. Moreover, the longer the storage time for the commercial product, the greater the maximum modulus value. In summary, the self-prepared formulation was consistent with the commercial product in terms of rheological behavior, and performed better in the linear viscoelastic region and uniformity, as well as possessing stronger resistance to deformation.