We designed a low-frequency (1.04 MHz) array with a large pitch (2
λ) for the experiments to achieve an array with a large aperture (30
λ) that can provide a large manipulation space and trapping force, thereby enabling
in vivo applications. The size of the controllable particles is related to the frequency of the array; based on the low-frequency array and size of the traps (the diameters of the focal spot and vortex centre were approximately 3 and 1 mm, respectively), we selected the PDMS and PS particles with a millimetre scale for manipulation. The results also demonstrated that the trapping force and Rayleigh limit can be improved by decreasing the pitch of the array; the pitch should be reduced to near or less than
if possible. However, when the aperture is invariant, a smaller pitch means more array elements, which imposes new challenges to the design and manufacture of arrays and the support of excitation systems. In this study, under the premise of obtaining sufficient trapping force and manipulating space, we selected a pitch of 2
λ to overcome the trade-off between the small pitch and the existing excitation systems and the manufacturing technology of the array. The proposed method of single-array acoustic manipulation was found to be superior compared to array-based acoustic tweezers that use the standing wave methods in terms of setup flexibility and cost reduction. Further, our method prevented the appearance of secondary nodes along the
-axis, which resolved a pressing problem in holographic acoustic tweezers [
31]. However, there were limitations to our method. Particle manipulation in the beam propagation direction was less accurate and stable because the axial restraint only depended on the balance of gravity, buoyancy, and acoustic radiation force, while the axial radiation force was exerted upwards instead of converging at the trapping position. The instability of motion was also evident in the heterogeneous medium, where the particles floated upwards and downwards in the axial direction during movement. The distances from the transducer plane to the stably trapped particles are related to the preset focal distance and the energy exerted on the particles. In this study, the preset focal depth from 4 cm to 12 cm (beam width 2.1 mm to 3.3 mm) was selected for trapping to reduce the impact of the grating lobes and to maintain the good focusing ability of the array (Figure S6). The relationships between the driving voltage, preset focal distance, and actual trapping position were explored with particles of different sizes (0.61
λ, 0.93
λ, and 1.41
λ), respectively. As shown in Figure S7, when the focal distance is constant, the trapped particles move upward with an increasing voltage. Once the driving voltage is increased to a certain extent, the particles begin to vibrate and cannot be trapped stably. By adjusting the combination of the preset focal distance and the driving voltage, the particles may be trapped more accurately at a specified height. The results also indicated that the maximum and the minimum distances from the transducer plane to the particles that can be trapped were 17.1 cm and 3.7 cm, respectively. The relationships among the duty factor, pulse repetition frequency (PRF), and position of the manipulated particle had also been studied. The duty factor used in this paper was 1.3%, and the PRF was 2.7 kHz. The influence of the duty factor on manipulation is mainly reflected in the energy that is exerted on the manipulated particles. When the PRF is constant, the particles are driven to move up with the increasing duty factor owing to the increased energy (Figure S8). However, when the duty factor is constant, increasing PRF has little influence on the trapping position (Figure S9).